collapse_allProperties

Top  Previous  Next

 

Boonton55xxx IVI Driver Reference


 

IBoonton55xxxAcquisition Properties




Name

Description

Public property

FetchLatency

The maximum amount of latency a fetch measurement may have. This setting adjusts how often automatic measurements and marker measurements are recalculated.

Public property

InitiateContinuous

Set or return the data acquisition mode for single or free-run measurements. If INITiate:CONTinuous is set to ON, the instrument immediately begins taking measurements (Modulated, CW and Statistical Modes), or arms its trigger and takes a measurement each time a trigger occurs (Pulse Mode). If set to OFF, the measurement will begin (or be armed) as soon as the INITiate command is issued, and will stop once the measurement criteria (averaging, filtering or sample count) has been satisfied. Note that INITiate:IMMediate and READ commands are invalid when INITiate:CONTinuous is set to ON; however, by convention this situation does not result in a SCPI error.

Public property

MeasurementMode

There are three basic measurement modes - pulse power, modulated power, and statistical power. Each mode is targeted towards a specific type of measurement.The Pulse Mode is used with peak power sensors. The instrument functions as an enhanced peak power analyzer and provides the functionality of a random repetitive sampling oscilloscope for viewing and analyzing the RF power envelope of signals.The Modulated Mode is designed for continuous, true average power measurements of complex modulated signals. The Statistical Mode is used for advanced analysis of non-periodic modulated signals. This mode does not require a trigger event to make measurements and is useful for signals that are noise-like such as CDMA, OFDM, or QAM. In Statistical Mode, the RF signal is sampled continuously at high speed, without discarding or losing any data.

Top

Footer image

Copyright 2013-16 Boonton. All rights reserved.