collapse_allSyntax

Top  Previous  Next

 

Boonton55xxx IVI Driver Reference


 

IBoonton55xxxAcquisition.Initiate Method


IBoonton55xxxAcquisition See Also



Starts a single measurement cycle when INITiate:CONTinuous is set to OFF. In Modulated Mode, the measurement will complete once the power has been integrated for the full FILTer time. In Pulse Mode, enough trace sweeps must be triggered to satisfy the AVERaging setting. In Statistical Mode, acquisition stops once the terminal condition(s) are met. In each case, no reading will be returned until the measurement is complete. This command is not valid when INITiate:CONTinuous is ON, however, by convention this situation does not result in a SCPI error

Namespace: Boonton.Boonton55xxx.Interop

Assembly: Boonton.Boonton55xxx.Interop (in Boonton.Boonton55xxx.Interop.dll)

Visual Basic

Public Sub Initiate ( _
)

C#

public void  Initiate(
)

Visual C++

HRESULT Initiate(
);

collapse_allInstrument Command

INITiate

collapse_allRequirements

Supported Models

All supported

collapse_allSee Also

Reference

IBoonton55xxxAcquisition Interface

Boonton.Boonton55xxx.Interop Namespace