Performs a calibration at a fixed frequency and level. 
Namespace: Boonton.Boonton55xxx.Interop 
Assembly: Boonton.Boonton55xxx.Interop (in Boonton.Boonton55xxx.Interop.dll)   
Parameters
Calibrator 
Thje calibrator type to use during fixed cal. 
Remarks
Fixed-cal does not provide for automatic control of the internal 50 MHz calibrator. The RF output level of the calibrator in use must be set by the user. Fixed-cal assumes that a valid Zero has already been performed. 
 Instrument Command
CALibration{Channel}:{INT|EXT}:FIXedcal 
 Requirements
Supported Models
All supported 
 See Also
Reference
IBoonton55xxxChannel Interface 
Boonton.Boonton55xxx.Interop Namespace 
  
Copyright 2013-16 Boonton. All rights reserved. 
 |