Set or return the fine position of the trigger event on displayed sweep. Namespace: Boonton.Boonton55xxx.Interop Assembly: Boonton.Boonton55xxx.Interop (in Boonton.Boonton55xxx.Interop.dll)
Property ValueSet or return the fine position of the trigger event on displayed sweep. RemarksThe position is given in divisions relative to the left edge of the screen, so with zero trigger delay, setting the vernier control to 0.0 causes the entire trace to be post-trigger. Setting it to 10.0 causes the entire trace to be pre-trigger. And setting to 5.0 will display both the pre- and post-trigger portions of the trace.
Note that the trigger delTRIGger:VERNierTRIGger:VERNierTRIGger:VERNierTRIGger:VERNay setting is in addition to this setting, and will cause the trigger position to appear in a different location. Setting the trigger vernier places the power meter in trigger vernier mode, which overrides the fixed "Left, Middle, Right" settings of the trigger position command. Instrument CommandTRIGger:VERNier {value} TRIGger:VERNier? Range-30 ≤ value ≤ 30 RequirementsSupported ModelsAll supported See AlsoReferenceIBoonton55xxxTrigger Interface |