Set or return the pulse amplitude percentage, which is used to define the end of a rising edge or beginning of a falling edge transition.
Namespace: Boonton.Boonton55xxx.Interop
Assembly: Boonton.Boonton55xxx.Interop (in Boonton.Boonton55xxx.Interop.dll)
Visual Basic
|
Public Property Distal As Single
|
C#
|
public float Distal { get; set; }
|
Visual C++
|
HRESULT get_Distal(
float* val
);
HRESULT put_Distal(
float val
);
|
Property Value
Set or return the pulse amplitude percentage, which is used to define the end of a rising edge or beginning of a falling edge transition.
Remarks
Typically, this is 90% voltage or 81% power relative to the top level of the pulse. This setting is used when making automatic pulse risetime and falltime calculations returned by READ:ARRay:AMEASure:POWer.
Units
Percent
Instrument Command
SENSe{Channel}:PULSe:DISTal {value}
SENSe{Channel}:PULSe:DISTal?
Range
50 ≤ value ≤ 100
Requirements
Supported Models
All supported