Set or return the units for entering the pulse distal, mesial and proximal levels. Attribute Tree Node: \Boonton55xxx\Instrument Specific\Channel\Pulse\Pulse Units Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. RepCapIdentifier A physical or virtual repeated capability identifier. For valid values, see the Channel repeated capability. AttributeID Must be BTN55XXX_ATTR_PULSE_UNITS. AttributeValue (GetAttribute) Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute. AttributeValue (SetAttribute) The value to which to set the attribute. RemarksIf the function is set to VOLTS, the pulse transition levels will be defined as the specified percentage in voltage. If set to WATTS, the levels are defined in percent power. Many pulse measurements call for 10% to 90% voltage (which equates to 1% to 81% power) for risetime and falltime measurements, and measure pulse widths from the half-power (-3dB, 50% power, or 71% voltage) points. Instrument CommandSENSe{Channel}:PULSe:UNIT {WATTS|VOLTS} SENSe{Channel}:PULSe:UNIT? Defined Values