Top Previous Next |
Set or return the trigger slope or polarity. When set to POS, trigger events will be generated when a signal?s rising edge crosses the trigger level threshold. When NEG, trigger events are generated on the falling edge of the pulse. Attribute Tree Node: \Boonton55xxx\Instrument Specific\Trigger\Edge\Slope Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. RepCapIdentifier Must be VI_NULL or an empty string. This attribute is not defined on a repeated capability. AttributeID Must be BTN55XXX_ATTR_SLOPE. AttributeValue (GetAttribute) Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute. AttributeValue (SetAttribute) The value to which to set the attribute.
|
Name |
Description |
BTN55XXX_VAL_TRIGGER_SLOPE_POSITIVE |
A positive (rising) edge passing through the trigger level triggers the meter. |
BTN55XXX_VAL_TRIGGER_SLOPE_NEGATIVE |
A negative (falling) edge passing through the trigger level triggers the meter. |