collapse_allSyntax

Top  Previous  Next

 

Boonton55xxx IVI Driver Reference


 

Btn55xxx_ReadIntervalMax Function


Interval See Also



Return the maximum instantaneous power or voltage in the time interval between marker1 and marker 2. The units will be the same as the specified channel.

Function Tree Node: \Boonton55xxx\Instrument Specific\Channel\MarkerMeasurements\Interval\Read Interval Max

Declaration: Btn55xxx.h

Implementation: Btn55xxx.dll

Visual C++

ViStatus Btn55xxx_ReadIntervalMax(
 ViSession Vi,
 ViConstString Channel,
 ViInt32ConditionCode,
 ViReal32Val
);

Parameters

Vi

The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session.

Channel

The physical or virtual repeated capability identifier. Pass VI_NULL or empty string if operation does not apply to a repeated capability.For valid values, see the Channel repeated capability.

ConditionCode

Condition Code

Val

Maximum instantaneous power or voltage in the time interval between marker1 and marker 2.

Return Value

 

Success or failure code.

collapse_allInstrument Command

READ{Channel}:INTERval:MAXimum?

collapse_allRequirements

Supported Models

All supported

collapse_allDefined Values For Parameter ConditionCode

Name

Description

BTN55XXX_VAL_COND_CODE_MEASUREMENT_STOPPED

-1 Measurement is STOPPED. Value returned is not updated.

BTN55XXX_VAL_COND_CODE_ERROR

Error return. Measurement is not valid.

BTN55XXX_VAL_COND_CODE_UNDERRANGE

An Under-range condition exists.

BTN55XXX_VAL_COND_CODE_OVERRANGE

An Over-range condition exists.

BTN55XXX_VAL_COND_CODE_NORMAL

1 Normal return. No error.