Btn55xxx_SetMode Function
|
|
Trigger See Also
|
|
Set or return the trigger mode for synchronizing data acquisition with pulsed signals.
Function Tree Node: \Boonton55xxx\Instrument Specific\Trigger\Set Mode
Declaration: Btn55xxx.h
Implementation: Btn55xxx.dll
Visual C++
|
ViStatus Btn55xxx_SetMode(
ViSession Vi,
ViInt32 Mode
);
|
Parameters
Vi
The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session.
Mode
Set or return the trigger mode for synchronizing data acquisition with pulsed signals.
Return Value
Success or failure code.
Instrument Command
TRIGger:MODE {NORMAL|AUTO|AUTOPKPK|FREERUN}
Requirements
Supported Models
All supported
Defined Values For Parameter Mode
Name
|
Description
|
BTN55XXX_VAL_TRIGGER_MODE_NORMAL
|
The power meter causes a sweep to be triggered each time the power level crosses the preset trigger level in the direction specified by the slope.
|
BTN55XXX_VAL_TRIGGER_MODE_AUTO
|
The power meter automatically triggers if the configured trigger does not occur within the meter's timeout period.
|
BTN55XXX_VAL_TRIGGER_MODE_AUTO_LEVEL
|
The power meter automatically adjusts the trigger level the trigger level to halfway between the highest and lowest power levels detected.
|
BTN55XXX_VAL_TRIGGER_MODE_FREERUN
|
The power meter forces traces at a high rate to assist in locating the signal.
|
Top
See Also
Reference
Trigger
Btn55xxx