Syntax |
Top Previous Next |
Set or return the pulse amplitude percentage, which is used to define the end of a rising edge or beginning of a falling edge transition. Attribute Tree Node: \Boonton55xxx\Instrument Specific\Channel\Pulse\Distal Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. RepCapIdentifier A physical or virtual repeated capability identifier. For valid values, see the Channel repeated capability. AttributeID Must be BTN55XXX_ATTR_DISTAL. AttributeValue (GetAttribute) Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute. AttributeValue (SetAttribute) The value to which to set the attribute. RemarksTypically, this is 90% voltage or 81% power relative to the top level of the pulse. This setting is used when making automatic pulse risetime and falltime calculations returned by READ:ARRay:AMEASure:POWer. Instrument CommandSENSe{Channel}:PULSe:DISTal {value} SENSe{Channel}:PULSe:DISTal? Range50 ≤ value ≤ 100 RequirementsSupported ModelsAll supported See AlsoReference |