Syntax |
Top Previous Next |
Set or return the data acquisition mode for single or free-run measurements. If INITiate:CONTinuous is set to ON, the instrument immediately begins taking measurements (Modulated, CW and Statistical Modes), or arms its trigger and takes a measurement each time a trigger occurs (Pulse Mode). If set to OFF, the measurement will begin (or be armed) as soon as the INITiate command is issued, and will stop once the measurement criteria (averaging, filtering or sample count) has been satisfied. Note that INITiate:IMMediate and READ commands are invalid when INITiate:CONTinuous is set to ON; however, by convention this situation does not result in a SCPI error. Attribute Tree Node: \Boonton55xxx\Instrument Specific\Acquisition\Initiate Continuous Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. RepCapIdentifier Must be VI_NULL or an empty string. This attribute is not defined on a repeated capability. AttributeID Must be BTN55XXX_ATTR_INITIATE_CONTINUOUS. AttributeValue (GetAttribute) Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute. AttributeValue (SetAttribute) The value to which to set the attribute. Instrument CommandINITiate:CONTinuous {} INITiate:CONTinuous? RequirementsSupported ModelsAll supported See AlsoReference |