collapse_allSyntax

Top  Previous  Next

 

Boonton55xxx IVI Driver Reference


 

BTN55XXX_ATTR_POSITION Attribute


Trigger See Also



Set or return the position of the trigger event on displayed sweep.

Attribute Tree Node: \Boonton55xxx\Instrument Specific\Trigger\Position

Declaration: Btn55xxx.h

Implementation: Btn55xxx.dll

Visual C++

#define BTN55XXX_ATTR_POSITION 1150021
ViStatus Btn55xxx_GetAttributeViInt32(
 ViSession Vi,
 ViConstString RepCapIdentifier,
 ViAttr AttributeID,
 ViInt32AttributeValue
);
ViStatus Btn55xxx_SetAttributeViInt32(
 ViSession Vi,
 ViConstString RepCapIdentifier,
 ViAttr AttributeID,
 ViInt32 AttributeValue
);

Parameters

Vi

The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session.

RepCapIdentifier

Must be VI_NULL or an empty string. This attribute is not defined on a repeated capability.

AttributeID

Must be BTN55XXX_ATTR_POSITION.

AttributeValue (GetAttribute)

Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute.

AttributeValue (SetAttribute)

The value to which to set the attribute.

collapse_allRemarks

Assuming zero trigger delay, setting the position to LEFT causes the entire trace to be post-trigger. Setting it to RIGHT causes the entire trace to be pre-trigger. And setting to MIDDLE will display both the pre- and post-trigger portions of the trace.

 

Note that the trigger delay setting is in addition to this setting, and will cause the trigger position to appear in a different location. Setting the trigger position places thepower meter in trigger position mode, which overrides the variable vernier settings of the trigger vernier command.

collapse_allInstrument Command

TRIGger:POSition {LEFT|MIDDLE|RIGHT}

TRIGger:POSition?

collapse_allDefined Values

Name

Description

BTN55XXX_VAL_TRIGGER_POSITION_LEFT

Assuming zero trigger delay, setting the position to Left causes the entire trace to be post-trigger.

BTN55XXX_VAL_TRIGGER_POSITION_MIDDLE

Setting trigger position to Middle will display both the pre- and post-trigger portions of the trace.

BTN55XXX_VAL_TRIGGER_POSITION_RIGHT

Assuming zero trigger delay, setting trigger position to right causes the entire trace to be pre-trigger.