Syntax |
Top Previous Next |
Set or return the position of the trigger event on displayed sweep. Attribute Tree Node: \Boonton55xxx\Instrument Specific\Trigger\Position Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. RepCapIdentifier Must be VI_NULL or an empty string. This attribute is not defined on a repeated capability. AttributeID Must be BTN55XXX_ATTR_POSITION. AttributeValue (GetAttribute) Returns the current value of the attribute. The user must specify the address of a variable that has the same data type as the attribute. AttributeValue (SetAttribute) The value to which to set the attribute. RemarksAssuming zero trigger delay, setting the position to LEFT causes the entire trace to be post-trigger. Setting it to RIGHT causes the entire trace to be pre-trigger. And setting to MIDDLE will display both the pre- and post-trigger portions of the trace.
Note that the trigger delay setting is in addition to this setting, and will cause the trigger position to appear in a different location. Setting the trigger position places thepower meter in trigger position mode, which overrides the variable vernier settings of the trigger vernier command. Instrument CommandTRIGger:POSition {LEFT|MIDDLE|RIGHT} TRIGger:POSition? Defined Values
RequirementsSupported ModelsAll supported See AlsoReference |