Performs a zero offset null adjustment. Function Tree Node: \Boonton55xxx\Instrument Specific\Channel\Zero Channel Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. Channel The physical or virtual repeated capability identifier. Pass VI_NULL or empty string if operation does not apply to a repeated capability.For valid values, see the Channel repeated capability. Calibrator The calibrator type to turn off during zeroing. Return Value
RemarksThe sensor does not need to be connected to any calibrator for zeroing ? the procedure is often performed in-system. However, this command will turn off the internal calibrator (when installed) prior to zeroing to avoid the need to perform this step explicitly. Instrument CommandCALibration{Channel}:{INT|EXT}:ZERO RequirementsSupported ModelsAll supported Defined Values For Parameter Calibrator
See AlsoReference |