Syntax |
Top Previous Next |
Performs a calibration at a fixed frequency and level. Function Tree Node: \Boonton55xxx\Instrument Specific\Channel\Fixed Cal Declaration: Btn55xxx.h Implementation: Btn55xxx.dll
ParametersVi The ViSession handle that you obtain from the IviDriver_init or IviDriver_InitWithOptions function. The handle identifies a particular instrument session. Channel The physical or virtual repeated capability identifier. Pass VI_NULL or empty string if operation does not apply to a repeated capability.For valid values, see the Channel repeated capability. Calibrator Thje calibrator type to use during fixed cal. Return Value
RemarksFixed-cal does not provide for automatic control of the internal 50 MHz calibrator. The RF output level of the calibrator in use must be set by the user. Fixed-cal assumes that a valid Zero has already been performed. Instrument CommandCALibration{Channel}:{INT|EXT}:FIXedcal RequirementsSupported ModelsAll supported Defined Values For Parameter Calibrator
See AlsoReference |